7.0

 

Recalibration 

Sensitivity and Zero Measurand Output calibration should be performed at 6 to 12 month intervals, depending on usage.  
Ordinarily, recalibration need be performed only at 12 month intervals if it is known that the angular rate sensor has not been 
used  beyond  its  rated  specifications.    If  the  unit  is  used  under  severe  environments,  it  may  be  desirable  to  use  shorter 
calibration intervals. 

Contact Endevco for local  calibration facility information or return the 7310A unit to Endevco  for recalibration.  Endevco 
maintains an angular rate recalibration service with A2LA traceability in the United States.  

Dirty units may be wiped clean using a damp cloth and a solvent such as acetone. 

8.0

 

Questions 

 

If you have any questions regarding the use of this or any Endevco sensors, please contact 

mailto:sales@endevco.com

  or 

go to our website 

https://endevco.com/

   for information of Endevco Application Engineering in North America and your 

local sales representatives. 

Summary of Contents for 7310A

Page 1: ...IM7310A Rev A April 7th 2020 7310A Angular Rate Sensor IM7310A Instruction manual ...

Page 2: ... and evaluations were conducted EMC standards EN 55011 Class A conducted and radiated RF emissions EN 61000 3 2 Harmonic current emissions EN 61000 3 3 Voltage fluctuations and flicker EN 61000 4 2 Electrostatic discharge immunity EN 61000 4 3 Radiated electromagnetic field immunity EN 61000 4 4 Fast transient burst immunity EN 61000 4 6 Conducted RF immunity EN 61000 4 11 Voltage dips short inter...

Page 3: ...Pre installation Check up 4 5 0 Mounting 4 5 1 Mounting Surface 5 5 2 Screw Mount 5 5 3 Adhesive Mount 6 5 4 Cable 6 5 5 Electrical Connection 6 5 6 Triaxial Mount 6 5 7 6DoF Mount 7 6 0 Electrical Precautions 8 6 1 Excitation Voltage 8 6 2 Signal Leads 8 6 3 Grounding 8 6 4 Signal Conditioning 8 7 0 Recalibration 9 8 0 Questions 9 ...

Page 4: ... of the angular rate sensor it should be checked to insure proper operation A simple Output Voltage test can be conducted with minimal test equipment Place the unit on the flat surface of a vibration free table top apply the specified excitation voltage to the angular rate sensor see data sheet for proper wiring hook up and measure the output with a voltage meter Allow the unit to warm up for one ...

Page 5: ...n wrench EHM2156 alloy steel black oxide finish as shown in Figure 1 Remove the unit from the shipping container Place the unit on the mounting surface and align the mounting holes Slide the washers over the screws Using the supplied wrench or a torque wrench tighten the screws to 4 lbf in 0 45 Nm This is roughly equivalent to finger tight with the supplied wrench The Model 7310A angular rate sens...

Page 6: ...Electrical Connection Connect the unit to the signal conditioner using the following lead designation Red Excitation Green Output White Output Black Ground Excitation Endevco Model 136 3 channel signal conditioner is an optional accessory and is recommended to be used with Model 7310A 5 6 Triaxial Mount For triaxial angular rate measurements the Model 7930 Triax Six Degrees of Freedom Mounting Blo...

Page 7: ...the three angular rate sensors should be installed to the three primary mounting surfaces marked of the 7930 mounting block and they can be mounted on either the top or the bottom surface of the sensor depending on the positive direction of rotation in the specific application The three accelerometers are installed on the other three surfaces of the mounting block measuring linear accelerations al...

Page 8: ... signal leads are differential The green lead has a varying output while the white lead is fixed at a reference voltage of approximately 2 5 Vdc Although the output is high level at 2 Vdc the usual precautions of using twisted pair shielded instrumentation cable should be taken to avoid noise pickup on the signal leads 6 3 Grounding The case of the model 7310A angular rate sensor is not attached t...

Page 9: ...use shorter calibration intervals Contact Endevco for local calibration facility information or return the 7310A unit to Endevco for recalibration Endevco maintains an angular rate recalibration service with A2LA traceability in the United States Dirty units may be wiped clean using a damp cloth and a solvent such as acetone 8 0 Questions If you have any questions regarding the use of this or any ...

Reviews: