Recalibration
Sensitivity and Zero Measurand Output calibration should be performed at 6 to 12 month intervals, depending on usage.
Ordinarily, recalibration need be performed only at 12 month intervals if it is known that the angular rate sensor has not been
used beyond its rated specifications. If the unit is used under severe environments, it may be desirable to use shorter
calibration intervals.
Contact Endevco for local calibration facility information or return the 7310A unit to Endevco for recalibration. Endevco
maintains an angular rate recalibration service with A2LA traceability in the United States.
Dirty units may be wiped clean using a damp cloth and a solvent such as acetone.