100
Rev B 2022/03
Troubleshooting
CyBio Well vario
Table 14: Further error messages
Text too long
Text which is intended for
output to the system display
is too long.
Shorten output text.
stage reached upper end
position
stage motion has reached
its upper end position.
Use keyboard at the CyBio
Well vario to trigger stage
motion into lowermost
position. To do this, press
both direction keys at the
same time. This will auto-
matically trigger a drive cal-
ibration routine.
stage reached lower end
position
stage motion has reached
its lower end position.
Use keyboard at the CyBio
Well vario to trigger stage
motion into lowermost
position. To do this, press
both direction keys at the
same time. This will auto-
matically trigger a drive cal-
ibration routine.
Unknown keyboard code
The code of a key which was
pressed at the system con-
trol panel is unknown to
software. This error occurs
following an attempt to
trigger a program file that
was created using a more
recent than the currently
installed software version.
Use latest software version
in this case. If this can be
discarded as potential error
cause, the particular file is
likely to have suffered dam-
age. Create a new program
file.
Volume too small
The sum of all volumes
aspirated and dispensed is
too small. Valid values are
within the range of 2
μ
l to
250
μ
l.
If all signed cycle values are
added up with each other,
the result must not be
below the lower limit of this
range.
Correct setting values
appropriately.
Capillary head does not
dispense in a conforming
manner.
Blocked or broken capillar-
ies
Replace clogged or broken
)
No air supply or faulty con-
troller setting of com-
pressed air control unit.
Check air supply. In the
event of controller misad-
justment at the compressed
air control unit, restore the
correct operating pressure
(
Capillary head does not
correctly dispense into the
microplate
Microplate mispositioned
Check configuration for cor-
rect plate type setting.
Error message
Cause
Remedy
TIP
Motion of all electric-powered mechanical components will immediately stop on open-
ing of the head cover.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...