Rev B 2022/03
129
CyBio Well vario
It may happen that after dabbing or even while dabbing on the wet fleece, a thin film of
liquid remains / forms on the tips. In this case, it is necessary to burst this film with an
increased additional volume/blowout. Since only a limited blowout volume is available
mechanically, 1
μ
l of air (air gap) must be taken up before pipetting in this case, which is
available as additional blowout volume at the end of the cycle (the 1
μ
l additional volume
taken up as air gap at the beginning of the procedure is not included in the script excerpt
shown below).
Fig. 134: Example 3
Move to dab location
Switch on vacuum
Move the dabbing station very close to
Residual ejection 1
μ
l
Waiting time
Remove tips from the fleece
Waiting time
Contact/dab again
Waiting time
Switch off vacuum
Lower lifter
Move pistons to zero position
Contact with fleece
fleece –
no
contact
!
Waiting time
Ejection of the nominal volume
(air gap / additional volume)
Remove tips from the fleece
Residual ejection with blowout
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...