CyBio Well vario
Technical Description
Designed for high precision, including in the submicro liter volume range, and with great
versatility, the CyBio Well vario covers a large range of applications. Through
adaptation of the positioning system in XY-motion direction, all basic functions of a 96-
channel simultaneously working system are also available for 384-channel microplates,
using a four-step sequence, and for 1536-channel microplates, using a sixteen-step
sequence. A 384-channel simultaneously working system will require four steps to com-
plete a 1536-well microplate.
On integration of one or two stackers, up to fifty microplates can be processed in a fully
automated sequence per stacker.
CyBio Well vario operation is controlled via CyBio Composer, a software facilitating easy
and expeditious editing of specific laboratory routines. In addition, manual commands may
be triggered from the operation control panel
(
4.5.2
Function With Capillary Head
The CyBio Well vario with capillary head allows for concurrent compound transfer into dry
microplates of small volume capacity (from 25 nl to 1000 nl). This works particularly well
where small amounts of liquid must be transferred quickly and in parallel, for example, in:
Compound management
Compound (dry) storage
Assay development, screening/HTS, ADME/Tox
Genomics, proteomics, ...
Dispensing stock is aspirated by capillary force. This means that 96 or 384 capillaries can
be filled simultaneously using the effect of capillary force. In principle, a sample transfer
consists of dipping the capillaries into the liquid of a source plate and blowing the contents
of the capillaries into a target plate.
Once the capillaries are submerged in the liquid of a source plate, capillary force aspirates
liquid into the capillaries so the capillaries fill up "of their own accord". Dispensing is
accomplished with a short jet of compressed air to transfer the contents of capillaries to a
target plate. System operation is also controlled by CyBio Composer software. Manual
operation is equally possible (
The active capillary wash station ensures reliable cleaning of the capillaries.
Basic functions
Explanation
Dispensing
• Aspirates a given total volume, making use of capillary force
effect
• Dispenses a given total volume with a jet of compressed air
Tip replacement
• Performs electro-motoric unclamping of a magazine plate with
96 or 384 capillaries
• Capillary magazine is removed from the system
• New capillary magazine is inserted
• Performs electro-motoric clamping and sealing of magazine
plate.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...