51
CyBio Well vario
Initial Start-Up & Routine Start-Up Procedure
6.2
Configuration & Start-Up
In view of the complexity and to ensure failsafe system operation, work for installation and
initial start-up of the CyBio Well vario on your premises is must be completely performed
by customer service personnel of the manufacturer or duly authorized expert technicians.
Initial start-up essentially includes:
Installation and adjustment of all system components
Providing cable connections and connecting power supply cables
Software installation (factory-performed) and configuration
Device induction
Check for integrity, completeness and compliance with packing list as you unpack the prod-
uct shipment. On completion of work for system installation, customer service personnel
will perform checks for proper function, providing documented proof of successful testing.
„Current firmware version” on page 9
.
6.3
Function Tests
6.3.1
Precision Test
Pipetting head CyBio Well vario 96/
384
Testing for variation coefficient (variation coefficient percentage standard deviation) is
performed, using a dye solution in a 96 well or 384 well microplate with flat transparent
bottom. A suitable vertical photometer is used for measuring. The photometer's own preci-
sion must be verified and documented according to manufacturer instructions prior to
measurement.
Materials & preparatory action:
96 well or 384 well transparent flat-bottom microplate prefilled with 0.1 N NaOH
Tabelle 9, “Prefill, aspiration and test volumes”, auf Seite 52
Table 8: Prefill volumes
Always use new tips for precision measurement. Rinse new tips before measurement
with p-nitrophenol working solution – settings:
Testing is performed in manual mode, using an aspiration volume of 20
μ
l or 10
μ
l,
respectively, and a dispensing volume of 10
μ
l, 3
μ
l or 2
μ
L, respectively.
Dispense the test volume into a microplate (see table below) with prefilled NaOH solu-
tion. For test volume dispensing, the Pipette tips should submerge into the prefilled
NaOH solution by approximately 1 mm.
Dispense the remaining volume into the storage reservoir.
CyBio Well vario
96-well/250
μ
L
CyBio Well vario
96-well
/60
μ
l/40
μ
l/25
μ
l
CyBio Well vario 384-
well
60
μ
l/25
μ
l
Diluens
150
μ
l
150
μ
l
50
μ
l
TIP
A certain inhomogeneity in the rate of evaporation over a microplate will influence the
result of testing adversely.
For this reason, you should tape the microplates immediately after diluens has been
filled in, after pipetting of the test volume and before shaking. Do not use washed mi-
croplates. This may cause excessively great variation in measured values.
Piston speed:
100 rpm
Number of rinse cycles:
20
Rinsing volume:
10
μ
l
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...