52
Rev B 2022/03
Initial Start-Up & Routine Start-Up Procedure
CyBio Well vario
*Select a dye concentration corresponding to an extinction between 0.4 and 1.2 OD
at 405 nm.
Table 9: Prefill, aspiration and test volumes
Tape the microplates after pipetting of the test volume and before shaking.
Use an orbital shaker for mixing of substances.
Table 10: Mixing times
Measure extinction in vertical photometer.
Evaluate measured data as a preparatory stage of CV value determination.
Compare obtained CV result with CV values in chapter
.
Pipetting head CyBio Well
vario 1536/8
μ
l
Testing for variation coefficient (variation coefficient percentage standard deviation) is
performed, using a dye solution in a 1536 well microplate with flat transparent bottom. A
suitable vertical photometer is used for measuring. The photometer's own precision must
be verified and documented according to manufacturer instructions prior to measurement.
CyBio Well vario
96-well/250
μ
L
CyBio Well vario
96-well/ 40
μ
l/
25
μ
l
CyBio Well vario
384-well/ 25
μ
l
CyBio Well vario
96/384-well/60
μ
l
Aspiration volume
20
μ
l
10
μ
l
10
μ
l
10
μ
l
Standard test vol-
ume*
10
μ
l
2
μ
l
2
μ
l
3
μ
l
p-nitrophenol
dye solution
125 mg/l in
0.1 N NaOH
600 mg/l in
0.1 N NaOH
200 mg/l in
0.1 N NaOH
125 mg/l in
0.1 N NaOH
96-well microplate
384-well microplate
Orbital shaker
with
700 rpm
Shake for 15 min
30min break
Shake for 15 min
—
Centrifuge with
2000 rpm
Centrifuge 2 min
15 min break
Centrifuge 2 min
Centrifuge 2 min
15 min break
Centrifuge 2 min
Orbital shaker
1
with
1100 rpm
—
Shake for 15 min
45 min break
Shake for 15 min
1 Alternative option (with orbital shaker)
TIP
Dispense residual fluid (in the tips) onto the dabbing station (if possible, "Pipetting"
mode).
In any case, the tips should be dabbed off on the dabbing station before completion of
the pipetting/dispensing cycle (Composer: piston in zero position; operation control
panel: dabbing break) in order to avoid carryover due to residual liquid that always re-
mains on the tips.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...