Rev B 2022/03
95
CyBio Well vario
Troubleshooting
8
Troubleshooting
Malfunction or faults during operation will be displayed on the working screen. Where a
fault situation was clearly caused by operator action, operation can be resumed, once the
fault has been removed (switch power off and on again for greater safety).
Check all potential fault sources on occurrence of a fault. If problems remain after this
check or on identification of an undescribed fault, you should notify the customer service of
the manufacturer or their authorized service partner.
8.1
Error Messages
The following fault messages may be displayed by the CyBio Composer software:
Error/error code
Cause
Remedy
1
Tip clamping device defective
Notify customer service
2
Undefined head type
Notify customer service
3
RAM fault
Notify customer service
4
Tips are loose
Switch power supply off and on
again.
5
Magazine not fully inserted
Remove and insert magazine
once again.
Tip magazine of older make/
back corners are not rounded.
Replace tip magazine or have
magazine reworked by dealer/
service personnel.
Tip magazine edges or corners
deformed resulting from drop
impact.
Replace tip magazine or have
deformation removed.
7
System error
Notify customer service
8
Failure to find internal zero-
position of pump system.
You may be able to remove this
error by performing an
exchange of tips.
Note:
On repeated occurrence of this
fault, repair work is necessary!
9
Parameter error: host computer
has transferred wrong parame-
ters to CyBio Well vario pipettor.
Check for system configuration
in CyBio Composer software.
10
Host computer triggered a com-
mand that is not available in the
pipettor.
Check command and repeat
command transfer.
11
Pump drive has reached upper
limit position.
You may be able to remove this
error by performing an
exchange of tips.
Note:
On repeated occurrence of this
fault, repair work is necessary!
12
Pump drive has reached lower
limit position.
You may be able to remove this
error by performing an
exchange of tips.
Note:
On repeated occurrence of this
fault, repair work is necessary!
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...