Rev B 2022/03
109
CyBio Well vario
Decommissioning
10 Decommissioning
If the CyBio Well vario is not required for a longer period of time, perform shut-down
action as described hereafter:
1.
Remove the tip magazine and place it into the tip tray package or remove the capillary
magazine and place it into its special storage box.
2.
Install plate adapter in the head.
3.
Trigger a head change routine (
) to remove the head from
the system.
4.
Only for capillary head:
Disconnect compressed air supply tube from connector of
compressed air control unit. Leave short tube end connected to the head.
Fig. 121: Disconnectible air supply inlet
5.
Transfer power switch into position "0".
6.
Only for capillary head:
Switch compressed air supply off and drain the system space
until pressure-relieved state is achieved.
7.
Switch system power supply off (using main switch or socket bar for this purpose).
8.
Disconnect power cord from socket bar or line power socket.
9.
Disconnect compressed air supply tube from inlet connector at compressed air control
unit.
10. Remove all microplates.
11. Clean and disinfect the pipettor and its components as described in handling instruc-
tions for the most recently used materials and substances.
12. Protect the system from sedimentation of dust.
The system has been completely shut down.
CAUTION
There is danger of injury and damage to the pipettor if cables are removed before pow-
er is switched off!
Do not remove cables as long as they are energized! Make absolutely certain that power
supply has been cut before you remove a cable!
TIP
A sub-menu you had been working in when power was switched off will be restored when
power becomes available again.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...