Technical Description
Fig. 29:
Typical behavior of the residual liquid in the case of
small pipette tips and capillaries
1 Tip in liquid after residual ejection
2 Tip is moved out of liquid, residual liquid film remains
3 Piston is moved to zero position; residual liquid is aspirated into pipette tip
4 Liquid film remains in the pipette tip
5 (and 6) Carryover in the next pipetting cycle
Proper use of the dabbing station can prevent this carryover.
For the accurate transfer of small volumes, reverse pipetting with additional stroke or addi-
tional volume is generally recommended in order to exclude falsification of the uptake vol-
umes due to capillary effects.
Residual ejection should always be performed on the dabbing station with
activated
vac-
uum in order to completely remove the residual liquid from the tips.
(
TIP
Before each resetting of the pistons to the zero position, it is
imperative
that the residual
liquid is completely removed from the tips by dabbing on the fleece surface of the dab-
bing station.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...