44
Rev B 2022/03
Technical Description
CyBio Well vario
After use, the dabbing station should be rinsed again with 30 to 50 ml of water.
In addition, rinse with 30-50 ml of 70 % ethanol to prevent bacteria, etc. from multiplying
in the moist environment.
The use of the dabbing station can also be advisable for normal tips, especially if the liq-
uids to be transferred tend to foam (e.g. contain proteins or detergents).
4.5
Function
4.5.1
Function With Pipetting Head
The CyBio Well vario multi-channel pipettor is designed as a 96-well or
384-well simultaneously working dosing system to allow automatic processing of micro-
plates in chemical, biological and physical research laboratories.
The system provides six basic functions which, on aggregate, offer a diversity of application
options (
). A pipetting head can also be quickly replaced with another
one (
). Available options are:
Table 7: System functions
Basic functions
Explanation
Dosing
• Aspirates a specified volume of liquid from a trough
• Dispenses previously aspirated liquid into a microplate (example)
with additional stroke
Pipetting
• Aspirates slightly more than a predefined volume of liquid from
trough (with additional stroke volume).
• Discharges predefined volume, leaving residual volume in pipette
tip (differential pipetting).
• Dispenses residual volume, for example, into a trough
Dispensing
• Aspirates a given total volume as the sum of all defined subvol-
umes
• Dispenses a specified number of subvolumes in a corresponding
number of steps
Diluting
• Aspirates a great specified volume
• Aspirates an air bubble
• Aspirates a small specified volume according to a required mixing
ratio
• Dispenses the total content of tips with additional stroke
Purging
• Aspirates and dispenses a specified volume of rinsing liquid in a
specified number of rinse cycles
Tip replacement
• Performs electro-motoric release of 96-tip, 384-tip or 1536-tip
magazine plate.
• Removes magazine plate from the system.
• Removes used pipette tips from magazine plate.
• Fills new set of tips into magazine plate.
• Inserts filled magazine plate.
• Performs electro-motoric clamping and sealing of magazine plate.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...