76
Rev B 2022/03
Operation
CyBio Well vario
Click
Allows you to set an acoustic click for confirmation of each keystroke. You can switch the
click on and off.
Capillary head
Pipettor settings can be adapted to specific user or application requirements. This can be
achieved via the
[Configuration]
menu:
1.
Restores the main menu should the pipettor be working in a different submenu.
2.
Use key
or
in the main menu to select the
[Configuration]
submenu.
3.
Confirm your selection with the
key.
You are working in the [Configuration] menu now.
.
Fig. 86:
Display – configuration menu (capillary head)
4.
Select appropriate parameter and enter desired values in
[Configuration]
menu,
(refer to description below). Confirm each input by pressing
.
Working in the
[Configuration]
menu you can set the following parameters:
Stage motion length
For defining a maximum available length of stage motion (lifter height). This setting pre-
vents mechanical collision of a vessel with tips. It also facilitates different applications, for
example, the picking up of liquid over a cell lawn or the pipetting onto the bottom of a
microplate from a preselected distance. The height position can be adjusted in steps of 0.1
mm.
dZ
For updating the vertical setting value for the lifter following necessary replacement of the
lifter (setting value inscribed on lifter bottom on shipment). This will automatically correct
vertical lifter adjustment so previously edited programs need not be changed in terms of
height coordinates.
Wells
This option can be used to select a mode for manual lifter motion adapted to a given one
out of various plate types (e.g. motion to positions of a 384 well microplate or a 1536 well
microplate). The correct plate type must be set in order to guarantee that a tip is located in
CAUTION
With excessively high settings for carriage motion or rotary stage motion, there may be
overflow of liquid from the reagent troughs. Select a reasonably reduced speed value if
necessary.
TIP
In manual mode, lifter motion is limited to a maximum speed of 17 mm/s (60 rpm). This
is necessary to enable a more precise height setting. In program-controlled (computer-
controlled) mode, the limiting value is around 40 mm/s (140 rpm).
Stage height
dZ
Wells
XY
Stage speed
Beep
Exit
CAUTION
There is danger of tips suffering damage or getting destroyed!
A value setting for lifter height will only be effective in manual mode so it may be ex-
ceeded in software-controlled operation mode!
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...