The solid plate edge must safely withstand the gripping force (25 N).
Fig. 12
Labware dimensions
Minimum length (l)
127.76 mm ± 0.5 mm
Minimum depth (d)
85.48 mm ± 0.5 mm
Typical height (h + h‘)
(h‘ = height of the plate edge)
14.35 mm ± 0.76 mm
The device transports the labware in landscape format.
3.4
Scope of delivery
The following components are included in the delivery:
¡
CyBio Carry plate transport system, consisting of:
– CyBio Carry X (linear axis) variant
– CyBio Carry Y/Z (gripper tower)
¡
External power supply unit and connecting cable, RS 232 cable
¡
Packaging box with transport locks
The following technical documentation is included in the scope of delivery:
¡
Declaration of incorporation
¡
Assembly instructions
¡
Software documentation (CyBio Composer)