CyBio Carry
Installation and commissioning
19
4
Installation and commissioning
4.1
Installation conditions
4.1.1
Ambient conditions
Observe the following location requirements:
¡
This laboratory device is designed for inside use.
¡
Do not use the device in explosion-hazard environments.
¡
Do not locate the device near sources of electromagnetic interference.
¡
Place the device on a stable, horizontal surface.
¡
Avoid mechanical shocks and vibrations.
¡
The installation site must be free of drafts, dust and caustic fumes.
¡
Do not use the device in wet and damp environments. Keep the device surface clean
and dry.
¡
Avoid direct sunlight and radiation from heaters onto the device. If necessary, pro-
vide air conditioning.
¡
Allow the device to acclimatize after transport and storage. Otherwise, condensed
air moisture could damage the sensitive electronics during commissioning.
¡
Position the device so that the emergency stop button and the on/off switch are al-
ways easily accessible.
The following requirements are placed on the climatic conditions in the operating room:
¡
Operating temperature: +15 to +35 °C
¡
Humidity during operation: ≤75 % at +35 °C, non-condensing
4.1.2
Spatial requirements
The required space depends on all components that make up the system.
The CyBio Carry plate transport system has the following dimensions (depending on the
model used):
Width
735 to 2375 mm (→ "Technical data"
Height
530 mm
Depth
¡
140 mm
¡
400 mm (with protective glass and maximally extended gripper
fingers)
Provide sufficient space for a control computer, monitor, and other PC accessories.
4.1.3
Power supply
WARNING
Danger due to electrical voltage
¡
Only connect the device to a properly grounded socket which complies with the volt-
age specified in these instructions (→ "Technical data"
46).
¡
Do not use an adapter in the feeder.