Maintenance and care
HS 55 (Batch Mode)
42
Figure 19Gold collector
1 Gas inlet
2 Heating wire
3 Gas outlet
4 Heating connection
5 Infrared sensor
6 Heating connection
}
Unscrew the gas inlet and gas outlet hoses from the gold collector.
}
Pull the heating coil connections off the PCB.
}
Loosen the gold collector screw fitting at the compartment. Remove the gold collec-
tor with heating coil and pull off the screw fitting.
}
Insert the new gold collector into the screw fitting.
}
Insert the gold collector into the compartment. Simultaneously insert the isolation
sleeves on the heating wire into the groove.
}
Slide the gold collector up to the stop and screw it tight.
}
Plug the connections of the new heating coil onto the PCB.
}
Tighten the gas hoses on the right and left side of the gold collector with hollow
screws.
ü
The new gold collector is ready for operation.
6.8
Maintaining the HydrEA system
6.8.1
Cleaning the coated graphite tube
The iridium or gold-coated graphite tube in the HydrEA system can be cleaned by baking
out. This procedure can also be repeated multiple times.
The iridium coat evaporates at temperatures above 2200 °C, the gold coat at more than
1000 °C. Do not exceed these temperatures when cleaning.
}
Start the ASpect LS or ASpect CS program.