Troubleshooting
HS 55 (Batch Mode)
44
7
Troubleshooting
Strong foaming can occur in the sample during the hydride and Hg cold vapor meth-
ods.
¡
Test the foaming of unknown samples.
¡
Immediately stop the measuring process if the transport gas argon transports
foam up to the quartz cell.
¡
Add a few drops of anti-foaming agent to strongly foaming samples, e.g., Dow-
Corning DB 110A, silicone anti-foaming agent or octanol.