5 Operation
ASP station 2000
74
EHauser
5.4
Confirmation of error messages
Confirmation of error
messages
Acknowledge error messages on the controller display by pressing the operating key
beneath the OK field.
5.5
Communication using ReadWin
®
2000
Communication using
ReadWin
®
2000
Besides by means of local operation, you can also remotely configure the device and
poll measured values using a PC with the Hauser operating software
ReadWin
®
2000. ReadWin
®
2000 is a universal PC programme for remote device
operation. The interface for remote configuration is the RS232 connection socket. You
can find detailed information on operating the device using ReadWin
®
2000 on the CD-
ROM with the device software supplied.
5.6
Calibration
Calibration
5.6.1
Tap calibration
You must calibrate the tap if:
•
The tap motor was replaced
•
The error message: <Tap calibration> appears in the display.
Carry out the calibration as follows:
Fig. 20:
Tap calibration
1.
Switch the device on (see Section 6.2).
2.
Under < SET -> SERVICE - CALIBRATION>, select
the item < DIST. TAB >.
3.
When you select “Start”, the tap turns and stops just
before the calibration position.
4.
At the controller, keep selecting <1 step> until the
arrow on the front side of the tap is located exactly in
the notch in the middle of the distribution pan (point
A). In the picklist, select the menu option <SAVE>.
5.
The tap is calibrated.