Operation
ASP Station 2000 RPS20B
32
Hauser
Ext. stop
The input signal stops any programs running. The programs are continued when
the signal disappears.
Event
The input signal triggers "event sampling". For example, the input signal may be
switched when a limit value is infringed. A separate bottle can be filled when an
event sample occurs.
Report
A message appears on the display showing the date and time (e.g. function error
of the flowmeter). The message must be acknowledged, the sampling program is
not interrupted.
Start Aut
The input signal starts the configured main program.
Stop Aut
The input signal ends the configured main program.
Switch
The input signal triggers a switch to the switching program.
For the described functions, a digital signal must be assigned to digital input 1, 2 or
Analog input
Inputs
Digital Input 2
Digital Input 3
Analog Input
Analog Input
no
0-20mA
Esc
â
á
<-'
signal
4-20mA
units
l/s
dec.point
XXX,X
range
100,0 l/s
Esc
â
á
<-'
:
:
:
:
Digital Input 1
4-20mA
0-1V
A0023635-EN
13
Menu structure
Signal
Output signal of the connected device: none, 0 to 1 V, 0 to 20 mA or 4 to 20 mA
Units
SI unit of the analog signal
Dec. point
Number of decimal places of the analog signal
Range
Maximum measuring range
For the described functions, an analog signal must be assigned to the analog input.