Address
Description
Works settings
210
Programme 1: (Select one from three possibilities)
- Time proportional sampling:
Samples are taken in even time cycles.
-
Quantity proportional sampling:
High flow = Many samples
Low flow = Few samples
Condition: That a flow meter is connected to the sampler.
-
Event controlled sampling:
So long as address 270 is not set to "Ext.Signal" an external signal (positive edge)
at the control input initiates an immediate sample.
Timed
211
Time proportional sampling: Set up time between each sample cycle.
15 min.
212
Quantity proportional sampling: Set up the quantity at which the sampler is to initiate a sample cycle.
-
213
Distribution tap changeover: Select if timed or number of samples released initiates a change to next bottle.
Timed
214
Set up time for distribution tap change to next bottle.
2 hours
215
Set up number of samples to be taken before distribution tap change to next bottle.
-
220-225
Set up programme 2 (see Aadresses 210-215).
See matrix
230-235
Set up programme 3"
"
240-245
Set up programme 4"
"
250-255
Set up programm 5"
"
260-265
Set up programme 6"
"
270
Programme change
(Select one from four possibilities)
Not active:
No programme change.
Time:
Programme change at preset times.
Quantity:
An external flow meter must be connected to the sampler.
Programme change is dependent on preset high and/or low flow limits
Return (reset) hysteresis is 1% of preset value.
Ext.Signal:
Programme change is initiated by an external signal at the "control input"
Return (reset) when signal is no longer active.
Condition: That addresses 210/220 and so on up to 260 are not set to "Event".
Not active
271
A = Selected active programme (addr. 010).
B = Changeover target programme.
-
272
Programme change
Set times for changeover and return (reset).
-
273
...The same every day or select:
Changeover day (select one day from Monday to Sunday) and
return (reset) day (select one day from Monday to Sunday).
-
274
Programme change dependent on flow.
Set up limit for programme change.
-
275
Bottle change on programme change:
Select yes or no
Bottle change on programme return:
Select yes or no.
-
276
Immediate sample cycle on programme change: Select yes or no
280
Yes
= Sample sequence end after filling last bottle. This is indicated with a sequence end message
No
= Continuous operation (distribution tap continues to turn. Plan bottle change as these may overfill !!!).
Yes
asp-station d 2
Operation and display
29