Service level
410
Update
Service date
Yes / no
411
Software
Name and
version
number
412
Processor
report
Abort counter
3 digit
last fault
4 digit
420
Sampler
running time
Hours
6 digit
421
Pump
running time
Hours
each 6
digit
resettable
and totaliser
422
Cooler
running time
Hours
each 6
digit
resettable
and totaliser
423
Sample
counter
Each 6 digit
resettable
and totaliser
424
Electrode 2
errors
Each 4 digit
resettable
425
Ack. without
cleaning
Each 4 digit
resettable
and totaliser
430
Sample test
Function in
steps
431
Distribution
tap test
Change in
steps or on 1
container
440
Select
analogue
calib. range
0...20 mA
4...20 mA
0...1V
0...10V
441
Connect 0%
value
0 mA or
4 mA or
0 V
442
Connect
100% value
20 mA or
1 V or
10 V
443
Accept
calibrated
values
450
Calibrate
temperature
input
active or
not active
451
Calibrate
0°C
Connect 1615
Ω
reference
resistor
452
Calibrate
50°C
Connect 2372
Ω
reference
resistor
453
Accept
calibrated
values
460
Unit test
active or
not active
461
Test outputs
In/out:
output 1...3
heating
cooling
pump
462
Test inputs
Off, impulse
input,
stop input,
programme
change
463
Test serial
channel
TTY:
Connect input
with output
464
Test
analogue
input
0% value
100% value
Result:
XXX,X %
465
Test
temperature
input
Connect
1615
Ω
2372
Ω
470
Activate
RESET
Yes / no
Last on:..
471
Activate
PRESET
Yes / no
Last on:..
32
Operation and display
asp-station d 2