Cerabar PMP23 IO-Link
Operation options
Hauser
19
7
Operation options
7.1
Operation with an operating menu
7.1.1
IO-Link
IO-Link information
IO-Link is a point-to-point connection for communication between the measuring device
and an IO-Link master. The measuring device features an IO-Link communication interface
type 2 with a second IO function on pin 4. This requires an IO-Link-compatible assembly
(IO-Link master) for operation. The IO-Link communication interface enables direct access
to the process and diagnostic data. It also provides the option of configuring the measuring
device while in operation.
Physical layer, the measuring devices supports the following features:
• IO-Link specification: version 1.1
• IO-Link Smart Sensor Profile 2nd Edition
• SIO mode: yes
• Speed: COM2; 38.4 kBaud
• Minimum cycle time: 2.5 msec.
• Process data width: 24 bit
• IO-Link data storage: yes
• Block configuration: no
IO-Link download
http://www.endress.com/download
• Select "Software" as the media type.
• Select "Device Driver" as the software type.
Select IO-Link (IODD).
• In the "Text Search" field enter the device name.
https://ioddfinder.io-link.com/
Search by
– Manufacturer
– Article number
– Product type
7.1.2
Structure of the operating menu
The menu structure has been implemented according to VDMA 24574-1 and
complemented by Hauser-specific menu items.
For an overview of the operating menu, see → 42