Cleanfit CUA451
Installation
Hauser
15
5.1.3
Installation instructions
Installation location
The assembly is designed for installation on vessels and pipes. Suitable nozzles must be
available for this. The minimum pipe diameter is DN 80.
‣
Before mounting the sensor, mount the assembly on the vessel or in the pipe.
The following diagram shows different installation positions in pipes, and indicates
whether they are permitted or not.
A0038661
9
Schematic of installation positions and orientations
‣
Ideally, the assembly should be mounted in an ascending pipe. Installation in a
horizontal pipe is also possible.
• When using reflective materials (e.g. stainless steel), the pipe diameter must be at least
100 mm (4"). An onsite calibration is recommended.
• Install the sensor in places with uniform flow conditions.
• Do not install the sensor in places where air may collect or foam bubbles form or where
suspended particles may settle.
• Avoid installation in the down pipe.
• Avoid fittings downstream from pressure reduction stages which can lead to outgassing.
Orientations
1
3
4
2
A0038660
10
Orientations, schematic
1
Sensor (see Accessories)
2
Transmitter
3
CUA451 retractable assembly
4
Direction of flow
The orientation depends on the sensor head. Pay attention to the Operating
Instructions for the relevant sensor.