Installation
Cleanfit CUA451
22
Hauser
3.
1
1
A0038433
Tighten the grub screw of the bayonet nut.
4. Connect the rinse chamber connection.
Fitting the sensor in the assembly
A0038445
1. Push up the hand lever as far as possible.
The ball valve is open.
2. Grease the retraction pipe.
This helps the retraction pipe move more smoothly in the direction of the
measuring position.
3. Push the retraction pipe in the direction of the measuring position as far as it will go.
The retraction pipe with the sensor is in the measuring position.
4. Hold the bayonet lock steady and close it.
5. Secure the retraction pipe with the securing screws.
6. Turn the entire top assembly part around its own axis until the sensor is in the right
position in relation to the medium flow.
5.2.4
Fitting the sensor holder onto another sensor
The universal sensor holder can subsequently be fitted onto another sensor.
The sensor holder has several sensor-specific grooves.
The desired sensor is aligned at the grooves.