Cleanfit CUA451
Installation
Hauser
23
Disassembling the sensor holder
Use the following tools to install the sensor:
• 2.5 mm Allen screw
• 6 mm Allen screw
1.
1
2
A0038431
Release the securing screws (item 1) and put them in a safe place within reach.
2. Turn the bayonet nut.
The bayonet lock (item 2) is released.
3.
1
2
A0038432
Taking the handles, pull out the retraction pipe (item 2) together with the sensor
holder (item 1) as far as it will go.
4. Close the ball valve! Push down the hand lever as far as possible (only possible in one
direction!).
When the ball valve is closed, the assembly is sealed from the process.
5. Connect a hose to the vent valve.
6. Vent the rinse chamber.