Operation
Cleanfit CUA451
28
Hauser
7
Operation
7.1
Adapting the device to the process conditions
7.1.1
From the service position to the measuring position
1. Check the rinse chamber connections to ensure they are closed.
2. Open the ball valve.
3. Push the retraction pipe in the direction of the process as far as it will go.
4. Lock the retraction pipe via the bayonet lock.
5. Secure with the securing screws.
The sensor holder is now secure and cannot move back accidentally.
6. At the handles, turn the top assembly part around its own axis in order to align the
sensor.
7.1.2
From the measuring position to the service position
1. Release the securing screws with an Allen key.
2. Open the bayonet lock.
3. Pull out the sensor holder as far as it will go (service position).
4. Close the ball valve.
5. Vent the rinse chamber.
6. Perform the necessary service work.