Technical data
Cleanfit CUA451
40
Hauser
11
Technical data
11.1 Environment
Ambient temperature
range
0 to 50 °C (32 to 122 °F)
11.2 Process
Medium temperature
0 to 85 °C (32 to 185 °F)
Medium pressure
Max. 10 bar (145 psi)
For manual insertion/retraction of the assembly, the medium pressure must not
exceed 2 bar (29 psi)! Also take the process conditions of the sensor used into
consideration!
Pressure/temperature
ratings
80
70
60
50
40
30
20
10
2
4
6
8
10
29
58
87
116
145
p
p
[psi] [bar]
176
158
140
122
104
86
68
50
T[°C]
T[°F]
A
85
185
A0038761
17
Pressure/temperature ratings
A
Range in which the assembly can be operated manually
11.3 Mechanical construction
Dimensions
→ Section "Installation"
Rinse connection nozzles
Connection options:
• 2 x ball valve with hose connection OD 9mm (see "Accessories"). (A ball valve is included
in the delivery for the assembly. On its own it acts as a vent cock.)
• Customer's own rinse connections with G1/8 external thread
• 2 x G1/8 (internal)
Vent cock
Ball valve with hose connection OD 9 mm
Weight