CleanFit P CPA 472
3 Installation
10
Hauser
3.2.2 Installation location/position
The CPA 472 assembly is designed for installation on tanks or pipes. Suitable connections must
be available.
!
Hinweis!
It is imperative to maintain an installation angle of min. 15°.
Abb. 3.5: Installation examples: The minimum installation angle is 15° to the horizontal.
!
Hinweis!
•
The assembly requires a minimum nominal width of DN 80 for installation in a pipe.
•
For thinner pipes please use a flow vessel. These are available as accessories (DN 25, PVDF
or DN 50, PP; for Order Numbers, see Accessories section).
Abb. 3.6: Flow vessel for CPA 472
left:
DN 25, PVDF
right:
DN 50, PP
"
Achtung!
Risk of frost damage!
If there is a risk of temperatures dropping below 0°C, the assembly and any air or water tubes
must be heated.