Ecograph T, RSG35
Appendix
Hauser
141
Totalizer
Navigation
Expert → Application → Maths → Maths x → Totalizer
Direct access code: 400014-00x
Examples: Maths 1: 400014-000; Maths 4: 400014-003
Description
Initial setting for the totalizer. Useful when continuing measurements recorded to date
with an (electro)-mechanical counter.
Only visible if the result is = counter, operating time from status or operating time from
total.
User entry
Number (max. 15 digits)
Factory setting
0
Formula editor
Enter the desired calculation formula.
Analog, digital or already active mathematics channels can be used.
Formula editor
Navigation
Expert → Application → Maths → Maths x → Formula
Direct access code: 400002-000
x = place holder for selected mathematics channel
A text field with the formula currently used appears. If the field is empty a formula
has not yet been defined for the mathematics channel.
Description
Individual channels can be mathematically linked and calculated with functions. The
mathematics channels calculated in this way are treated as "real" channels, regardless of
whether they are connected conventionally or via fieldbus. Enter the desired calculation
formula.
Analog and digital channels can be used, as can mathematics channels that are already
active.
A formula with up to 200 characters can be created using this editor. If the formula is
finished, click OK to close the editor and accept the formula entered. The common entry
and arithmetic operators and inputs are described in detail in the following sections.
Inputs
Inputs are described in the formula using the following syntax:
Input type (signal type;channel number)
Type of input
Description
AI
Analog inputs
DI
Digital inputs
MI
Mathematics inputs