Appendix
Ecograph T, RSG35
94
Hauser
Options
Switched off, On after 10 min, On after 30 min, On after 60 min, Switched daily, Control
input
Factory setting
Switched off
This setting has no effect if the screen saver is controlled by a digital input→ 115
ON daily from
Navigation
Expert → System → Screen saver → ON daily from
Direct access code: 160001-000
Description
Set time (hh:mm) as of when the screensaver should be switched on (e.g. shift end time).
The screen saver is switched off as soon as the device is operated via onsite
operation. After 1 min of inactivity it switches back on automatically.
Only visible if screen saver = switched daily
User entry
Time (hh:mm)
Factory setting
20:00
OFF daily from
Navigation
Expert → System → Screen saver → OFF daily from
Direct access code: 160002-000
Description
Set time (hh:mm) as of when the screensaver should be switched off (e.g. shift start time).
Only visible if screen saver = switched daily.
User entry
Time (hh:mm)
Factory setting
07:00
Alarm response
Navigation
Expert → System → Screen saver → Alarm response
Direct access code: 160003-000
Description
"Off on alarm": If limit value violations occur or status signal "Maintenance required (Mxxx)"
or "Check function (Cxxx)" is active, the screensaver will be automatically deactivated.
"Always on": If limit value violations occur or status signal "Maintenance required (Mxxx)"
or "Check function (Cxxx)" is active, the screensaver will not be deactivated.
Active messages that require acknowledgment or an active status signal "Failure
(Fxxx)" or "Out of specification (Sxxx)" deactivate the screensaver always.
Options
Off on alarm, Always on