Field Xpert
6 Device Xpert software
Hauser
25
Using the FXA520
• Start the Device Xpert configuration software, select "View" and then "Configuration".
• The "Public Fieldgate" has already been set up. Click and hold the Public Fieldgate icon until
a pop-up window appears. Select "Properties" here.
Fig. 6-5: Screenshot - FXA520 configuration
• If you wish, give the connection a new name.
Please ensure that "Fieldgate FXA520" has been selected.
• The address range can be selected in accordance with the field device.
If you do not know the address, we recommend the setting 0 to 15.
Fig. 6-6: Screenshot - FXA520 Properties
• In the second tab "Communications", now enter host, port, username and password. This
information is available from your Fieldgate (see fig. 5-8).