Proline Promag 51
Troubleshooting
Hauser
69
9.3
Process error messages
!
Note!
Also observe the information on
44.
692
S: SIM. MEASURAND
!: # 692
Simulation of a measured variable active (e.g. mass flow). Switch off simulation.
Custody transfer mode
Simulation not possible
698
S: DEV. TEST ACT.
!: # 698
The measuring device is being tested on site via the test
and simulation device.
Custody transfer mode
Checking via a test device is not possible.
No.
Error message / Type
Cause
Remedy (spare parts
ä
73ff)
No.
Error message / Type
Cause
Remedy (spare parts
P = Process error
$
= Fault message (with an effect on the outputs)
! = Notice message (without an effect on the outputs)
401
P: EMPTY PIPE
$
: # 401
Measuring tube partially filled or empty
1.
Check the process conditions of the plant.
2.
Fill the measuring tube.
Custody transfer mode
Reset error message via the status input or via the "reset key".
461
P: ADJ. NOT OK
!: # 461
EPD adjustment not possible because the fluid's
conductivity is either too low or too high.
The EPD function cannot be used with fluids of this nature.
463
FULL = EMPTY
$
: # 463
The EPD adjustment values for empty pipe and full pipe
are identical, therefore incorrect.
Repeat adjustment, making sure procedure is correct
Custody transfer mode
In custody transfer mode, an EPD/OED adjustment is not
possible.