Proline Promass 500
Use in protective systems
Hauser
35
2. Compare current at output 1 with this default value.
Comparing the current values
The current values can be compared using one of the following methods:
• Measure the current of the DUT at the logic subsystem (process control system or safety-
related PLC).
• Measure the current at the DUT using an external, traceably-calibrated ammeter.
‣
Compare the current values.
Connecting the measuring equipment and external testing
• Connecting the measuring equipment in the measuring circuit
• External check of the passive current output
Requirements for the measuring equipment:
• DC current measuring uncertainty ±0.2 %
• DC current resolution10 µA
2
1
A
3
4
A0034446
5
External verification taking the example of a passive current output
1
Automation system with current input (e.. g. PLC)
2
Power supply unit
3
Ammeter
4
Transmitter
1. Connect the ammeter to the transmitter by looping it in series into the circuit.
2. Connect the power supply unit.
Evaluation of results - Part 2: Testing of current output 1
The amount of deviation between the measured current and the set point must not exceed the
measured error specified for the safety function. The deviation should not exceed ±1 % /
±300 µA.
‣
Note data relating to measured error→ 18.
Connecting the test
1. Re-activate the locked SIL mode→ 21.
2. Deactivate bypassing of safety function of process control system.
Summary of Contents for HART Proline Promass 500
Page 2: ......
Page 8: ...Certificate Proline Promass 500 8 Endress Hauser 2 Certificate A0033748 ...
Page 50: ......
Page 51: ......
Page 52: ...www addresses endress com ...