Proline Promass 500
Use in protective systems
Hauser
43
Evaluation of results - Part 2: Testing of current output 1
The amount of deviation between the measured current and the set point must not exceed the
measured error specified for the safety function. The deviation should not exceed ±1 % /
±300 µA.
‣
Note data relating to measured error→ 18.
Connecting the test
1. Re-activate the locked SIL mode→ 21.
2. Deactivate bypassing of safety function of process control system.
3. Document results of proof test in accordance with the safety management guidelines
applicable to the system.
NOTICE
At least 99 % of dangerous, undetected failures are detected using these test sequences
(PTC = 0.99). The influence of systematic errors on the safety function is not fully
covered by the test. Systematic faults can be caused, for example, by medium properties,
operating conditions, build-up or corrosion.
‣
If one of the test criteria from the test sequences described above is not fulfilled, the device
may no longer be used as part of a protective system.
‣
Take measures to reduce systematic errors.
Detailed information on:
• Orientation
• Medium properties
• Operating conditions
Operating Instructions→ 11
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Current output
1 to n simulation
–
Switch the simulation of
the current output on and
off.
On
Off
Value current
output 1 to n
In the
Current output
1 to n simulation
parameter, the
On
option
is selected.
Enter the current value for
simulation.
• 1. Default
value: Select
4.0 mA.
• 2. Default
value: Select
20.0 mA.
3.59 to 22.5 mA
3.59 mA