Life cycle
Proline Promass 500
44
Hauser
6.3.7
Heartbeat Technology
Heartbeat Technology continuously diagnoses whether failures have occurred. The scope of
the diagnostics in the SIL mode corresponds to the SFF.
Heartbeat Technology also allows operators to create documented proof that diagnostic
checks have been carried out and thereby supports the documentation of proof testing in
accordance with IEC 61511-1, Section 16.3.3, "Documentation of proof testing and
inspections".
NOTICE
The SIL mode needs to be disabled temporarily in order to perform heartbeat
verification.
‣
On completion of the verification, the SIL mode must be enabled again .
The
Heartbeat Verification
application package is available as an order option and can
be retrofitted on all measuring devices.
Please contact your Hauser service or sales organization to retrofit the device.
7
Life cycle
7.1
Requirements for the personnel
The personnel for installation, commissioning, diagnostics and maintenance must fulfill the
following requirements:
‣
Trained, qualified specialists must have a relevant qualification for this specific function
and task.
‣
Are authorized by the plant owner/operator.
‣
Are familiar with federal/national regulations.
‣
Before starting work, read and understand the instructions in the manual and
supplementary documentation as well as the certificates (depending on the application).
‣
Follow instructions and comply with basic conditions.
The operating personnel must fulfill the following requirements:
‣
Are instructed and authorized according to the requirements of the task by the facility's
owner-operator.
‣
Follow the instructions in this manual.