Wiring
ISEmax CAS40D
12
Hauser
Electrode identification
4.3
Post-connection check
Electrode
Color of membrane ring and
marking on screw head
1)
1)
Color codes as per IEC 757
Cable identification
Ammonium
RD
1, 2 or 3
Nitrate
BU
Potassium
YE
Chloride
GN
pH, ORP (incl. reference)
no marking
R
Device status and specifications
Notes
Is the exterior of the sensor, assembly, junction box or cable externally
damaged?
Visual check
Electrical connection
Notes
Does the supply voltage of the transmitter match the data on the nameplate?
Are the installed cables strain-relieved and not twisted?
Is the cable type route completely isolated on site?
Power cables / signal lines
Are all screw terminals tightened?
Tighten
Are all cable entries mounted, tightened and sealed?
In the case of lateral cable entries: Cable
loops facing downwards to allow water to
drip off.
Are all cable entries facing downwards or mounted laterally?
Are the power supply and signal lines connected correctly?