ISEmax compressed air cleaning unit (230 V AC)
Commissioning
Hauser
13
Connect the cleaning unit to the transmitter:
‣
PE
L1
N
PE
T2.0A
230V AC / 50Hz
C
44
M
x
AOR/2R/4R
42
41
1
A0016124
Connect voltage L1 to terminals 41 and 42 of an additional module (AOR, 2R or 4R).
6.2
Post-connection check
Device status and specifications
Notes
Are the sensor and cable free from damage on the outside?
Visual inspection
Electrical connection
Notes
Does the supply voltage of the connected transmitter match the data on the
nameplate?
Visual inspection
Are the installed cables strain-relieved and not twisted?
Is the cable type route completely isolated on site?
Power cables / signal lines
Are all cable entries installed, tightened and sealed?
In the case of lateral cable entries:
Cable loops facing downwards to
allow water to drip off.
Are all cable entries facing downwards or mounted laterally?
7
Commissioning
7.1
Function check
L
CAUTION
Avoid risk of injury due to incorrect power supply and incorrect or non-secure hose
connections!
‣
Check that all connections have been established correctly.
‣
In particular, check all hose connections to ensure they are secure and liquid cannot escape.
‣
Ensure that the supply voltage matches the voltage indicated on the nameplate.