Levelflex M FMP43 with PROFIBUS PA
Commissioning
Hauser
67
Maximum measured error
Typical values under reference operating conditions:
DIN EN 61298-2, percentage values with reference to the span.
If the reference conditions are not met, the offset/zero point resulting from the installation setup
may be up to ±12 mm. This additional offset/zero point can be compensated for by entering a
correction ("offset" function (057)) during commissioning.
Differing from this, the following measuring error is present in the vicinity of the upper and lower
probe end:
Output:
Digital
Analog
Sum of non-linearity,
non-repeatability
and hysteresis
±3 mm
± 0.06 %
Offset / zero point
±4 mm
± 0.03 %
L00-FMP43xxx-05-00-00-en-001
3 mm
0 mm
- 3 mm
100 mm
200 mm
80 mm
200 mm
40 mm 20 mm
10 mm
- 10 mm
20 mm
- 20 mm
30 mm
- 30 mm
40 mm
- 40 mm
50 mm
- 50 mm
DK > 7
DK > 7
60 mm
- 60 mm
DK = 2
DK = 2
reference point of measurement
probe
end
maximum
measured
error
probe
end
(water-based
medium)
(water-based
medium)
(oils and solvents)
(
)
oils and solvents