Commissioning

Liquiline Mobile CML18

26

Hauser

7.4  

Setting the display language

1. Navigate to: 

Display language

Main menu

 >> 

System/Language

 >> 

Display language

2. Press   to scroll through the predefined values.

Description of setting

Configuration options

Change the language of the operating menu.

• Deutsch
• English

7.5  

Configuring the measuring device

7.5.1  

Configuring the Bluetooth connection

1. Navigate to: 

Bluetooth

Main menu

 >> 

System/Language

 >> 

Bluetooth

2. Press   to scroll through the predefined values.

Description of setting

Configuration options

Switch Bluetooth connection on/off

• Enabled
• Disabled

If the Bluetooth connection is disabled, operation via the SmartBlue App is not possible.

7.5.2  

Setting date and time

Preparatory steps

1. Enable Bluetooth. →   26
2. Link the device to a mobile terminal via the SmartBlue app. →   20

1. Select the device in the SmartBlue App.
2. Select 

System

.

3. Select 

date/time

.

4. Select 

Take over from mobile device

.

 or:

5. Manually configure the date and time.

7.6  

Advanced settings

7.6.1  

Displaying device information

1. Navigate to: 

Device info

Main menu

 >> 

Diagnostics

 >> 

Device info

Summary of Contents for Liquiline Mobile CML18

Page 1: ...Products Solutions Services Operating Instructions Liquiline Mobile CML18 Multiparameter mobile device BA02002C 07 EN 05 22 00 71558219 2022 01 26 ...

Page 2: ......

Page 3: ...tions 16 6 1 Overview of operation options 16 6 2 Internal operating menu with keys 16 6 3 Operation via SmartBlue app 20 7 Commissioning 25 7 1 Preliminaries 25 7 2 Function check 25 7 3 Switching on the device 25 7 4 Setting the display language 26 7 5 Configuring the measuring device 26 7 6 Advanced settings 26 8 Operation 32 8 1 Calibration 32 8 2 Reading measured values 32 9 Firmware update 4...

Page 4: ...gerous situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This symbol alerts you to a dangerous situation Failure to avoid this situation can result in minor or more serious injuries NOTICE Cause situation If necessary Consequences of non compliance if ap...

Page 5: ...Liquiline Mobile CML18 About this document Endress Hauser 5 1 3 Symbols on the device Symbol Meaning Reference to device documentation ...

Page 6: ...anufacturer s site or by the service organization The battery may only be changed directly at the manufacturer s premises or by the service organization 2 2 Intended use The Liquiline Mobile CML18 is a multiparameter mobile device for the connection of digital sensors with Memosens technology and optional operation by smartphone or other mobile devices via Bluetooth The device is designed for use ...

Page 7: ...otect them against unintentional operation 4 Label damaged products as defective During operation If faults cannot be rectified products must be taken out of service and protected against unintentional operation 2 5 Product safety 2 5 1 State of the art technology The product is designed to meet state of the art safety requirements has been tested and left the factory in a condition in which it is...

Page 8: ...auser 3 Product description 3 1 Product design 1 2 3 5 6 7 8 4 A0040968 1 CML18 1 Protection cap 2 Display screen with automatic screen rotation 3 Select button 4 Next button 5 Memosens connection 6 Area for wireless charging 7 Status LED 8 M12 connection ...

Page 9: ...ug in head and fixed cable sensors with the Memosens protocol and no external power supply pH ORP pH ORP combination sensors Conductive conductivity Inductive conductivity Dissolved oxygen optical amperometric In addition to measuring the main parameters Memosens sensors can be used to measure temperature The measuring range is adapted to the individual sensor type ...

Page 10: ...n in such a way that it is protected against impact and moisture The original packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact your supplier or your local Sales Center 4 2 Product identification 4 2 1 Nameplate The nameplate contains the following information Manufacturer identification Device designation Order ...

Page 11: ...ns All of the information relating to your device is displayed in this window as well as the product documentation Manufacturer s address Endress Hauser Conducta GmbH Co KG Dieselstraße 24 D 70839 Gerlingen 4 3 Scope of delivery The scope of delivery comprises 1 Liquiline Mobile CML18 1 set of Operating Instructions in German 1 set of Operating Instructions in English Inductive charger and power u...

Page 12: ...he Memosens sensor directly 1 2 A0040973 2 Sensor connection 1 Insert the sensor into the Memosens connection 2 Click the Memosens connection into place 5 1 2 Connecting the Memosens sensor with M12 fixed cable connection 1 A0040974 2 3 A0040972 1 Remove the protection cap 2 Insert the M12 fixed cable 3 Screw on the M12 fixed cable ...

Page 13: ...sens sensor 1 A0040974 3 Cable connection 2 3 A0040972 4 5 A0041005 1 Remove the protection cap 2 Insert the M12 connector 3 Screw on the M12 connector 4 Insert the sensor into the Memosens connection 5 Click the Memosens connection into place 5 2 Charging the device Charge the device fully before initial commissioning There are two possible ways to load the device wirelessly via Qi certified char...

Page 14: ...e melody sounds When device is switched off The green LED flashes during charging When charging is complete the charging complete melody sounds and the LED is continuously lit green for 10 minutes The device then switches off 5 2 1 Charging via Qi charger Only use Qi certified chargers Qi version 1 2 Further information www wirelesspowerconsortium com 1 2 A0044052 4 Inductive charging 1 Connect th...

Page 15: ...0972 4 A0044054 1 Remove the protective cap 2 Attach the M12 connector of the cable to the device connection 3 Screw on the M12 connector of the cable 4 Connect the USB connector to a USB charger or USB port on a computer 5 3 Ensuring the degree of protection Only the mechanical and electrical connections which are described in these instructions and which are necessary for the required designated...

Page 16: ... 1 Display and operating elements 3 2 1 A0040996 5 Overview of display and operating elements 1 Display 2 Select button 3 Next button Button functions Button Device switched off On measuring screen In the menu Switching on Scroll through measuring screens Scroll down Switching on Save current measured values Grab Sample Confirm select long hold Open the menu Switch to previous menu level measuring...

Page 17: ... Cond unit Res unit Erase data Erase grab values Abort Erase Erase continuous logs Abort Erase Data logger plot Units Diagnostics Sensor info Calibration info Diagnostics list Data logger entries Display test Device info Manufacturer Software version Serial number Description Extended order code System Language Display language Bluetooth Display brightness Signal sounds M12 CSV ...

Page 18: ...save w o charger Power off w charger Power off w o charger Regulatory information Support links Support links Guidance 1 point calib ORP 2 point calibration pH and ISFET Cell constant inductive conductive conductivity Installation factor conductive conductivity Air 100 rh oxygen Air variable oxygen 1 point calib oxygen ...

Page 19: ... The device and sensor are working reliably F Failure of device or sensor F status signal as per NAMUR NE107 M Device or sensor requires maintenance M status signal as per NAMUR NE107 C Device or sensor undergoing function check C status signal as per NAMUR NE107 S Device or sensor being operated out of specification S status as per NAMUR NE107 Structure of the measurement window The measurement w...

Page 20: ...via SmartBlue app The SmartBlue App is available for download from the Google Play Store for Android devices and from the Apple App Store for iOS devices Download the SmartBlue App Use the QR codes to download the app A0033202 7 Download links System requirements iOS devices iPhone 4S or higher from iOS9 0 iPad2 or higher from iOS9 0 iPod Touch 5th generation or higher from iOS9 0 Devices with And...

Page 21: ... be enabled on both devices Enable Bluetooth 26 A0044142 9 SmartBlue App Livelist The Livelist displays all of the devices that are within range Tap the device to select it To be able to use the device with the SmartBlue App the Bluetooth connection must be confirmed by entering a user name and password ...

Page 22: ...e admin 2 Initial password device serial number Change the user name and password after logging in for the first time The current measured values are displayed in the Home view The device information device tag serial number firmware version order code is also displayed ...

Page 23: ...3 2 4 A0048102 10 Home view of SmartBlue App with current measured values 1 CML18 system and device information 2 Shortcut to diagnostic list 3 Overview of measured values of connected sensor 4 General information and sampling option Operation is via 4 main menus ...

Page 24: ...contained sequence of activities e g for calibration Wizard guided operation Diagnostics Contains information on operation diagnostics and troubleshooting as well as configuration of the diagnostic behavior Application Sensor data for specific optimization and for detailed process adjustment Adapts the measuring point to the application System These menus contain parameters for configuring the ove...

Page 25: ...y if you can answer yes to all the following questions Device condition and specifications Are the device and all the cables free from damage on the outside Are the mounted cables strain relieved Are the cables routed without loops and cross overs 7 3 Switching on the device A0040976 12 Switching on the device Press or The device starts up A connected sensor is recognized automatically The time re...

Page 26: ...uetooth 2 Press to scroll through the predefined values Description of setting Configuration options Switch Bluetooth connection on off Enabled Disabled If the Bluetooth connection is disabled operation via the SmartBlue App is not possible 7 5 2 Setting date and time Preparatory steps 1 Enable Bluetooth 26 2 Link the device to a mobile terminal via the SmartBlue app 20 1 Select the device in the ...

Page 27: ...settings selected 1 Navigate to Power management Main menu System Language Power management 2 Press to scroll through the predefined values The following energy settings are available Power save w charger Power save w o charger Power off w charger Power off w o charger The power save mode is activated after the set time if there is no user interaction In the power save mode the display is switched...

Page 28: ... are 2 power off settings Power off w charger Functional description Configuration options Set the time until the device switches off automatically if it is connected to the mains 1 min 5 min 15 min 30 min 1 h 2 h Never Power off w o charger Functional description Configuration options Set the time until the device switches off automatically if it is running on the battery 1 min 5 min 15 min 30 mi...

Page 29: ...ain menu System Language M12 CSV 2 Press to scroll through the predefined values Description of setting Configuration options Switch on off M12 CSV On Off When the M12 CSV option is enabled no sensor can be operated via cable Operation via the Memosens connection on the device is still possible A message to this effect is shown on the display 7 6 5 Adjusting the display brightness 1 Navigate to Di...

Page 30: ...t measured value is saved automatically 1 s 2 s 10 s 20 s 30 s 1 min 5 min 30 min 1 h If the device is woken up to record a log value any existing switch on settling times of the connected sensor are not taken into account When measuring with oxygen sensors such as the Oxymax COS51D or COS22D the device with activated data logger remains permanently switched on regardless of the energy settings se...

Page 31: ...e data logger the measured value units can be adjusted for conductivity measurement with the data logger in ultrapure water An adjustment is necessary to eliminate rounding errors in the smallest measured values The units for conductivity and resistance can be permanently configured 1 Navigate to Cond unit Main menu Application Data logger Data logger Cond unit 2 Press to scroll through the predef...

Page 32: ...he desired calibration 2 Press to navigate through the calibration The following calibrations can be performed Type of calibration Measuring parameters Navigate to 1 point calibration ORP 1 point calib 2 point calibration pH or ISFET 2 point calibration Cell constant calibration Inductive conductive conductivity Cell constant Installation factor calibration Conductive conductivity Installation fac...

Page 33: ...reparatory steps 1 Enable Bluetooth 26 2 Link the device to a mobile terminal via the SmartBlue app 20 Transferring the data 1 Select the device in the SmartBlue App 2 Select Grab sample 3 Select ID text Click in the line of text to assign an individual text for the selected ID Depending on the input language selected there are up to 32 characters available for the assignment of the individual ID ...

Page 34: ...steps 1 Install the SmartBlue app on a mobile terminal 20 2 Enable Bluetooth 26 3 Link the device to a mobile terminal via the SmartBlue app 20 Transferring the data 1 Select the device in the SmartBlue app 2 Select in the SmartBlue app 3 A0042257 3 Select Guidance ...

Page 35: ...Liquiline Mobile CML18 Operation Endress Hauser 35 4 A0042258 4 Select Data transfer 5 A0042261 5 Continue with Next ...

Page 36: ...ecords of data logger 7 Press Ok to confirm Press to discard changes and close the drop down menu 8 Select File name Click the text line to enter an individual name for the generated data package 9 Press Ok to confirm Press to discard changes and close the drop down menu 10 Press Next to continue Data transfer starts A progress bar indicates the progress percentage ...

Page 37: ...Liquiline Mobile CML18 Operation Endress Hauser 37 11 A0042263 11 When the transfer is complete press Next to continue The result of the data transfer is displayed 13 12 A0042265 ...

Page 38: ...as Microsoft Excel The export files of the grab sample and data logger measured values have a different display format Elements of the export file Data logger export file Grab sample export file Elements in the general information section of the export file Filename File content Format version Device type Device tag Device serial number Device firmware version Sensor serial number PV name PV unit ...

Page 39: ...nique entry number This value is increased for each logged entry It is reset if the entries are deleted Status NAMUR device status when the entry is logged PV name Name of the primary value PV value Numerical display of the primary value of the logged entry PV unit Unit of the primary value SV name Name of the secondary value SV value Numerical display of the secondary value of the logged entry SV...

Page 40: ...n be updated to the latest version via the Smartblue App All saved data logger entries must be exported before each firmware update A firmware update can take up to one hour There must be sufficient battery charge if necessary connect the device to the mains 13 The device is prevented from switching off automatically if is connected to the SmartBlue app NOTICE Damage to firmware Risk of incomplete...

Page 41: ... 20 Starting a firmware update 1 Select the device in the SmartBlue App 2 Select in the SmartBlue app 3 Select System 4 Select Firmware update 5 Search for the available firmware update package on the terminal device and select it If the update is not displayed the firmware update package must be opened once using the SmartBlue App 6 Start the update 7 After updating the firmware successfully upda...

Page 42: ...n 10 1 2 Accessing calibration information 1 Navigate to Calibration info Main menu Diagnostics Calibration info 2 Press to access the calibration information 10 1 3 Opening the diagnostics list 1 Navigate to Diagnostics list Main menu Diagnostics Diagnostics list 2 Press to open the diagnostics list 10 1 4 Display testing 1 Navigate to Display test Main menu Diagnostics Display test 2 Press to in...

Page 43: ...t use organic cleaners such as acetone benzyl alcohol methanol methylene chloride xylene or concentrated glycerol cleaner for cleaning Do not use high pressure steam for cleaning 11 2 Measuring and test equipment Calibrated and adjusted sensors with Memosens technology save their calibration data directly in the sensor The sensors can be used as testing equipment thanks to this functionality The d...

Page 44: ... returning devices 12 2 Disposal The device contains electronic components The product must be disposed of as electronic waste Observe the local regulations If required by the Directive 2012 19 EU on waste electrical and electronic equipment WEEE the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste Do not dispose of products bearing t...

Page 45: ...3 1 M12 USB data charging cable Order code 71496600 Charging via cable Data backup Live data transfer A0047709 13 2 Protective cover Order code 71530939 Comprehensive protection Extremely robust Tabs and eyelets provide a range of securing options A0047710 ...

Page 46: ...CYK20 for sensors with Memosens technology A complete list of supported sensors is provided on the device product page www endress com CML18 Documents Manuals Software Certificates Supported sensors from the lab portfolio include CPL51E CPL53E CPL57E CPL59E CLL47E COL37E Supported sensors from the process portfolio include CPS11D CPS12D CPS16D CPS31D CPS41D CPS42D CPS47D CPS71D CPS72D CPS76D CPS77...

Page 47: ...V IEC 61 000 4 5 with 2 0 kV 14 3 5 Sensor connection Sensors with Memosens technology 14 3 6 Cable specification Digital measuring cable CYK10 Axx2 x Digital measuring cable CYK20 AAxxC1 M12 USB data charging cable 14 4 Environment 14 4 1 Ambient temperature range Charging 0 to 45 C 32 to 113 F Operation 10 to 60 C 14 to 140 F The maximum ambient temperature depends on the process temperature and...

Page 48: ... 48 Endress Hauser 14 4 5 Electrical safety EN 61010 1 14 4 6 Pollution degree Complete device Pollution level 4 Internal Pollution level 2 14 5 Mechanical construction 14 5 1 Dimensions 54 2 13 48 1 89 86 3 38 A0044044 14 Dimensions mm in ...

Page 49: ...907 2006 Art 33 1 The device battery contains the SVHC 1 3 propane sulton ethylene glycol dimethyl ether CAS number 1 110 71 4 with more than 0 1 w w The product does not present a hazard if it is used as designated 14 5 4 Impact loads The product is designed for mechanical impact loads of 1 J IK06 as per the requirements of EN 61010 1 14 5 5 Weight Liquiline Mobile CML18 155 g 5 5 oz 1 CAS Chemic...

Page 50: ... version 26 Dimensions 48 Display language 26 E Electrical connection 12 Electrical safety 48 F Firmware update 40 H Hardware reset 29 I Impact loads 49 Incoming acceptance 10 Input Measured variables 46 Intended use 6 L Language 26 M Manufacturer s address 11 Materials 49 Measured variables 46 Measuring parameters 9 Measuring range 46 N Nameplate 10 O Operation 32 Calibration 32 Device operation ...

Page 51: ...ta logger 33 Scope of delivery 11 Sensor Connection 47 Settings 26 Audio 28 Data logger 30 Display brightness 29 Energy settings 27 Signal sounds 28 Switching units 31 State of the art technology 7 Storage temperature 47 Supply voltage 47 Switching off 40 Switching on 25 Symbols 4 5 T Technical data 46 Environment 47 Input 46 Mechanical construction 48 Output 46 Technical personnel 6 Types of inpu...

Page 52: ...www addresses endress com 71558219 71558219 ...