10.7 Additional functions
10.7.1 Limit switches
There are different ways of configuring a limit switch:
• Assigning a switch-on and switch-off point
• Assigning a switch-on and switch-off delay for a relay
• Setting an alarm threshold and also outputting an error message
• Starting a cleaning function (for sensors)
Path: Menu/Setup/Additional functions/Limit switches/Limit switches 1 to 8
Function
Options
Info
Source of data
Options
• None
• Connected sensors
• Controllers available
• Analyzer
Factory setting
None
Specify the input or output which should be the source of
data for the limit switch.
Measuring value
Options
• None
• Depends on theSource of
data
Factory setting
None
The measured value you can select depends on the
option selected under "Source of data".
Measured valuedepending on theSource of data
Source of data
Measured value
Ortho-P
Options
• Main value
pH Glass
Options
• Raw value mV
• pH
• Temperature
pH ISFET
ORP
Options
• Temperature
• ORP mV
• ORP %