Diagnostics and troubleshooting
Liquiline System CA80PH
150
Hauser
728 Level cleaner
M;
On
Off
Level monitoring warning
Measurement can still take place.
Low cleaner level.
-> Top up cleaner and reset counter in
Diagnostics/Operating time information
729 Filter candle change
M;
On
Off
Filter cartridge needs replacement
The limit value for operating hours has been
exceeded
-> Replace filter cartridge of the sample
preparation system and reset the counter
730 Cleaning solution
M
On
Off
Level warning for cleaner of sample preparation
system
The remaining amount will last for a few hours or
days depending on the cleaning duration, the
cleaning interval and external events
1. Top up cleaner for the sample preparation
system
2. Check the level switch for the cleaning
solution of the sample preparation system
731 Leackage sensor
F
On
Off
Leakage detected in the sample preparation
system
1. Check hoses and connections
2. Check solenoid valves
3. Check leakage sensor
4. Replace defective parts and, if necessary,
reset the operating hours counters in the
Diagnostics menu
732 Wear parts alarm
F
On
On
One or more wear parts has exceeded its service
life.
-> Replace the wear parts in question and reset
the counter in Diagnostics/Operating time
information
733 Wear parts warning
M
On
Off
One or more wear parts has almost reached the
end of its service life.
-> Replace the wear parts in question and reset
the counter in Diagnostics/Operating time
information
910 Limit switch
S
On
Off
Limit switch activated
930 No sample
F
On
On
Sample flow interrupted during aspiration
• Suction line blocked or leaking
• No inflow of sample
-> Check suction line and suction strainer
-> Check sample flow
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current