Technical data
Liquiline System CA80PH
192
Hauser
15
Technical data
15.1 Input
15.1.1 Measured variables
PO
4
, PO
4
-P, P
2
O
5
[mg/l, ppm]
15.1.2 Measuring range
15.1.3 Types of input
• 1 or 2 measuring channels (main parameter analyzer)
• 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
• Analog current inputs (optional)
15.1.4 Input signal
Depending on version
• 2 x 0/4 to 20 mA (optional), passive, potentially isolated
15.1.5 Current input, passive
Span
> 0 to 20 mA
Signal characteristic
Linear
Internal resistance
Non-linear
Test voltage
500 V
15.1.6 Hose specification (self-priming)
• Distance: max. 1.0 m (3.3 ft)
• Height: max. 0.5 m (1.6 ft)
• Hose ID: 1.6 mm (1/16 inch)
CA80PH-AAE1:
0.05 to 2.5 mg/l PO
4
-P
(blue method)
CA80PH-AAE2:
0.05 to 10 mg/l PO
4
-P
(blue method)
CA80PH-AAE3:
0.5 to 20 mg/l PO
4
-P
(yellow method)
CA80PH-AAE4:
0.5 to 50 mg/l PO
4
-P
(yellow method)