15.5.3 Web server
The web server enables full access to the device configuration, measured values, diagnostic
messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G router with a
user-defined IP address.
15.5.4 EtherNet/IP
TCP port
80
Supported features
• Remote-controlled device configuration (1 session)
• Save/restore device configuration
• Logbook export (file formats: CSV, FDM)
• Access to web server via DTM or Internet Explorer
• Login
• Web server can be switched off
Protocol
EtherNet/IP
ODVA certification
yes
Device profile
Generic device (product type: 0x2B)
Manufacturer ID
0x049E
h
Device type identification
0x109F
Polarity
Auto-MIDI-X
Connections (max.)
CIP
12
I/O
6
Explicit message
6
Multicast
3 consumers
Minimum RPI
100 ms (default)
Maximum RPI
10000 ms
System integration
EtherNet/IP
EDS
Rockwell
Add-on profile Level 3
IO data
Input (T
→
O)
Device status and highest-priority diagnostic mes-
sage
Measured values:
• 16 AI (analog input) + unit
• 8 DI (discrete input) + status
Output (O
→
T)
Control values:
• 4 A0 (analog output) + unit
• 8 DO (discrete output) + status