Technical data
Liquiline System CA80PH
200
Hauser
15.7.8 Reagent requirement
• Blue method: approx. 50
μ
l per reagent and measurement
4)
• Yellow method: approx. 115
μ
l per measurement
15.7.9 Calibration interval
1 to 90 days, depending on the application and ambient conditions
15.7.10 Cleaning interval
1 to 90 days, depending on the application
15.7.11 Maintenance interval
Every 3 - 6 months, depending on the application
15.7.12 Servicing requirements
• Weekly: visual inspection
• Quarterly: 1 hour
15.8 Environment
15.8.1 Ambient temperature range
5 to 40 °C (41 to 104 °F)
15.8.2 Storage temperature
-20 to 60 °C (-4 to 140 °F)
15.8.3 Humidity
10 to 95%, not condensing
15.8.4 Degree of protection
IP55 (cabinet, analyzer stand)
15.8.5 Electromagnetic compatibility
Interference emission and interference immunity as per EN 61326-1: 2013, class A for industry
15.8.6 Electrical safety
IEC 61010-1, Class I equipment
Low voltage: overvoltage category II
Environment < 2000 m (< 6562 ft) above MSL
4)
The actual reagent shelf life can be shorter than the reagent longevity depending on the ambient conditions and dis-
play value