Electrical connection
Hauser
29
6.2
Connecting the analyzer
NOTICE
The device does not have a mains switch.
‣
You must install the device near (distance < 3 m (10 ft)) an easily accessible and fused plug
socket.
Depending on the order code, the mounted cable glands with an M-thread should be
replaced by the G' or NPT glands enclosed.
This does not affect the M32 hose glands.
6.2.1
Routing the cables
• Remove the bottle tray by lifting up the recessed grip slightly and then pulling it towards the
front.
• Remove the cover that is hooked into place.
• Using an Allen key, release the six screws on the carrier board and fold it out towards the
front. For convenience of handling, hook the carrier board onto the locking plate.
• Using a Phillips head screwdriver, release the six screws on the electronics compartment cover
and fold out the cover towards the front.
• Route the cables on the rear panel of the device so that they are properly protected.
• Cable glands are available for the cable entry.
• For cabinet versions, the cable length is approx. 4.3 m (14.1 ft) from the base of the housing.
For analyzer stand versions, the cable length is approx. 3.5 m (11.5 ft) from the base.
The terminal connection is located under an additional protective cover in the upper
section of the device.
6.2.2
Cable types
• Analog, signaling and transmission cables: e.g. LiYY 10 x 0.34 mm
2
The terminal cross-section must be at least 2.5 mm
2
for devices with 24V power supply.
With 24V power supply, a current of up to 10A can flow. For this reason pay attention to
the voltage drop on the supply line.The voltage at the device terminals must be within the
specified range (see "Supply voltage" section).