Electrical connection
Liquiline System CA80PH
46
Hauser
6.8.2
Bus address
Setting the bus address
‣
Set the desired bus address via the DIP switches of module 485.
anything between 1 and 247 for Modbus. If you configure an invalid address, software
addressing is automatically enabled via the local configuration or via the fieldbus.
6.9
Guaranteeing the degree of protection
Only the mechanical and electrical connections that are described in this manual, and are
necessary for the required, designated application, may be established on the device supplied.
‣
Please play close attention when performing the work as degrees of protection individually
confirmed for this product (ingress protection (IP), electrical safety, EMC interference
immunity) can no longer be guaranteed in the case of the following, for example:
– Leaving off covers
– Loose or insufficiently tightened cables/cable ends
a0016323
Fig. 59: Valid Modbus address 195
a0016324
Fig. 60: Invalid address 255
1)
1)
order configuration, software addressing is enabled, software address
configured at the factory: Modbus 247