Liquiline System CA80PH
Electrical connection
Hauser
47
6.10 Post-connection check
WARNING
!
Wiring errors
Incorrect wiring puts the safety of people and the measuring point at risk. The manufacturer
does not accept any responsibility for errors that result from failure to comply with the
instructions in this manual.
‣
Put the analyzer into operation only if you can answer
yes
to
all
of the following questions.
Device state and specifications
1.
Are the analyzer and cables free from damage on the outside?
2.
Are the hoses free from damage on the outside?
Electrical connection
3.
Are the mounted cables strain relieved?
Visual inspection of the liquid-bearing lines
4.
Is the suction line connected to the sample collector (if present)?
5.
Are the dispensers inserted correctly?
6.
Can the dispensers move up and down freely?
7.
Are all the hose connections leak-tight?
8.
If sample preparation is provided: has the connection been made? Are the protective hoses
in the hose glands strain relieved?
9.
If sample preparation is not connected, is the hose in the hose gland strain-relieved?
10. Have the bottles with reagents, cleaner and standard been inserted and connected?
11. Inspect the hose connections. Use the hose connection diagram as a guide.