System integration
Liquiline System CA80PH
50
Hauser
7.2
Service interface
You can connect the device to a computer via the service interface and configure it using
"Fieldcare". Furthermore, configurations can also be saved, transferred and documented.
7.2.1
Connection
‣
Connect the service interface on the controller housing to the Commubox (FXA291).
‣
Via the USB port, connect the Commubox to the computer.
‣
A CD is supplied with the Commubox and contains USB drivers that must be installed when
the device is connected for the first time.
‣
FieldCare must be installed on the computer. The software is either available on a CD or can
be downloaded from www.endress.com.
a0015904
Fig. 63: Connection overview
Additional information can be found in TI405C (FXA291) and CP010S (FieldCare) on the
CD.
7.2.2
Creating the data connection
1. Start Fieldcare.
2. Establish a connection to the Commubox.
3. To this end, for "Host PC" click "Add device" and select "CDI Communication FXA291".
4. Click "Configuration" to select the serial port for the FXA291 and to set the baud rate
(115200).
5. Click "Link connection" to start communication with the FXA291 (
are lit green).
6. Click "Add device" again and select your device type.
7. Click "Link connection" to start communication with the device (
are lit green).
8. To be able to communicate with the device online, select "Online parameterize".