Liquiline System CA80PH
System integration
Hauser
51
You can now start online configuration via the DTM.
Online configuration competes with onsite operation, i.e. each of the two options blocks the
other one. On both sides it is possible to take away access from the other side.
7.2.3
Operation
• In the DTM the menu structure corresponds to the onsite operation. The functions of the
Liquiline System soft keys are found in the main window on the left. This is where you also
navigate through the software.
• Clicking a menu name or a function corresponds to pressing the navigator.
• You can make your settings conveniently via the computer keyboard.
• Via Fieldcare, you can save logbooks, make backups of configurations and transfer the
configurations to other devices.
• Click "Operating devices", "Device functions" and "Further functions" to start the logbook export.
Select ".dat" in order to save the file in a protected format to process it with the Field Data
Manager.
• You can also print out configurations or save them as PDFs.
7.3
Fieldbuses
7.3.1
Modbus
With fieldbus module 485 and the appropriate device version, you can communicate via Modbus
RS485 or Modbus TCP.
The RTU and ASCII protocols are available when connecting via Modbus RS485. You can switch
to ASCII on the device.
Connect the Modbus data cable to the terminals of the fieldbus module (RS 485) or to the RJ45
(TCP) interface as described (
Further information on Modbus communication can be found on the product pages on the
internet (
→
SD01189C).
7.3.2
EtherNet/IP
With the fieldbus module 485 or module ETH and the appropriate device version, you can
communicate via EtherNet/IP.
Connect the EtherNet/IP data cable to the RJ45 jack on module 485 or ETH.
Further information on EtherNet/IP communication can be found on the product pages on
the internet (
→
SD01293C).