Operation
Liquiline System CA80PH
78
Hauser
Modbus
Diag. output
Options
• None
Factory setting
None
Before you can assign the message to an output, you
must first configure a relay output for "Diagnostics"
(Menu/Setup/Outputs, assign "Diagnostics" function and
set Operating modeto "as assigned" ).
Cleaning program (for
sensors)
Options
• None
• Cleaning 1
• Cleaning 2
• Cleaning 3
• Cleaning 4
Factory setting
None
Decide whether the diagnostic message should trigger a
cleaning program.
You can define the cleaning programs under:
Menu/Setup/Additional functions/Cleaning.
Detail information
Read only
Here you can find more information on the diagnostic
message and instructions on how to resolve the
problem.
Path: Menu/Setup/General settings/Extended setup/Modbus
Function
Options
Info
Enable
Options
• Off
• On
Factory setting
Off
You can switch off Modbus communication at this point.
The software can then only be accessed via local
operation.
Termination
Read only
If the device is the last in the bus, you can terminate via
the hardware.
--> "Electrical connection" section
Settings
Transmission Mode
Options
• TCP
• RTU
• ASCII
The transmission mode is displayed depending on the
version ordered. In the case of RS485 transmission, you
can choose between "RTU" and "ASCII".
Watchdog
0 to 999 s
Factory setting
5 s
If no data transmission takes place for longer than the
time set, this is an indicator that communication has
been interrupted.
After this time, input values received via the Modbus are
considered to be invalid.
Path: ... /Extended setup/Diagnostics settings/Diag. behavior
Function
Options
Info