Operation
Liquiline System CA80PH
92
Hauser
If Start condition
Date/time
is selected:
Date
(CAT860)
01.01.1970 to 07.02.2106
DD.MM.YYYY
Factory setting
01.01.1970
Time
(CAT860)
00:00:00 to 23:59:59
HH:MM:SS
Factory setting
00:00:00
Residence time
(CAT860)
00:30 to 20:00 (MM:SS)
Factory setting
01:00
Reaction time of cleaner during the cleaning
Sampling pump interval
(CAT820, CAT860)
Pump time
10 to 20 s
Factory setting
10 s
Ratio between the pumping duration and the pumping
interval of the peristaltic pump. This affects the volume
of the pumped sample.
Pump pause
20 to 50 s
Factory setting
30
Pressurized air cleaning
(CAT820, CAT860)
Compressed air
(CAT820)
Options
• Available
• Not available
Factory setting
Depends on the device
version
Settings for sample preparation with or without
compressed air cleaning.
In the event of retrofitting, the compressed air can be
activated here.
Cleaning mode
Options
• On
• Off
Factory setting
On
Activates or deactivates automatic compressed air
cleaning of the pump-to-filter hose and of the filter.
Cleaning interval
00:30 to 04:00 HH:MM
Factory setting
02:00
Cleaning interval of the automatic compressed air
cleaning system
Cleaning duration
10 to 60 s
Factory setting
30 s
Cleaning duration of the automatic compressed air
cleaning system
Path: Menu/Setup/Sample preparation
Function
Options
Info