Operation
Liquiline System CA80PH
96
Hauser
10.5 Binary inputs and outputs
Hardware options, e.g. "DIO" module with 2 digital inputs and 2 digital outputs or fieldbus
module "485" enables the following:
• via a digital input signal
– measuring range switching for conductivity (upgrade code required, see accessories)
– switching between different calibration data records in the case of optical sensors
– an external hold (for sensors)
– a cleaning interval to be triggered (for sensors)
– a PID controller to be switched on and off, e.g. via proximity switch
– the use of the input as an "analog input" for pulse-frequency modulation (PFM)
• via a digital output signal
– the static transmission (similar to a relay) of diagnostic states, point level switch states etc.
– the dynamic transmission (comparable to a non-wearing "analog output") of PFM signals,
e.g. to control dosing pumps.